| Marchio: | ACEY |
| Numero di modello: | ACEY-HRCDS-6V200A-32CH |
| MOQ: | 1 set |
| Prezzo: | USD 2480~23600/Set |
| Condizioni di pagamento: | T/T,PayPal,L/C,Western Union,MoneyGram |
| Capacità di approvvigionamento: | 200 insiemi/mese |
6V 300A 16 Channels Battery Life Cycle Tester For Li-ion/Lead-acid/Ni-Cd/Ni-MH Battery
Product Description
ACEY-HRCDS-6V300A-16CH battery life cycle tester is used for aging and life-cycle evaluation as well as quality inspection during battery manufacturing and laboratory testing. It is compatible with Lithium-ion, Lead-acid, Nickel-Cadmium (Ni-Cd), and Nickel-Metal Hydride (Ni-MH) batteries, and supports a full range of tests including cycle life, capacity, charge and discharge behavior, charge retention, energy efficiency, overcharge and over-discharge tolerance, pulse testing, and real-world operating condition simulations.
Equipment Configuration
| No. | Category | Product name | Product model | Quantity | Remarks |
| 1 | Test equipment | Charge and discharge test equipment | ACEY-HRCDS-6V300A | 16CH | |
| 2 | Cabinet | Charging and discharging equipment cabinet | / | 1PCS | Default without power input line |
| 3 | Computer | Computer | Computer I5-12500 processor /16G memory / 256G solid state + 1T hard disk / 3 network port | 1PCS | Optional,including mouse, keyboard and 21.5 inch displayer |
| 4 | Accessories | Charge discharge test line | / | 16PCS | Each channel has a set of test lines with a length of 3.5meters.(including fixture) |
| 5 | Accessories | Switch | TP-LINK TL-SF1008VE | 1PCS | |
| 6 | Accessories | Communication accessories | Communication line / Crystal head, etc | Several | For communication |
| 7 | Accessories | USB Flash Drive | 16G | 1PCS | Contains Test software etc. |
| 8 | Accessories | File | / | 1PCS | Test report, packing list, etc |
Equipment Features
Main Technical Parameter
| No. | Item | ACEY-HRCD6V300A-16CH | ||
| 1 | Main channels | 4CH/ module | ||
| 2 | Channel working mode | Completely independent. | ||
| 3 | Main channel working mode | Charge: CC, CV, CCCV, no voltage or current impact when CC transfer to CV. | ||
| Discharge: CD, CP, CR, DCCV. | ||||
| Still, Pulse test, Working Condition test, DCIR test. | ||||
| Main channel test cutoff condition | Time, total voltage, total current, capacity, -ΔV, Vcell, RSOC, etc. | |||
| 4 | Main channel Voltage | Single channel test range | -6V~6V | |
| Min. discharge voltage | -6V | |||
| Accuracy | ≤0.03%FS(@25℃±10℃),≤0.1%FS(@0-15℃,35-45℃) | |||
| Resolution | 1mV | |||
| Main channel Current | Single channel test range | ±300A | ||
| Min. output current | 50mA | |||
| Accuracy(automatic switching) | 50mA--50A:0.03%FS(@25℃±10℃) | |||
| 50A--200A:0.03%FS(@25℃±10℃) | ||||
| 200A--300A:0.03%FS(@25℃±10℃) | ||||
| Resolution | 1mA | |||
| Current start response time | <5ms(10%FS~90%FS) | |||
| Time | Resolution | 10ms | ||
| Auxiliary temperature (Optional) | Range | ﹣20℃~100℃ | ||
| Accuracy | ±1℃(0℃~70℃) | |||
| Resolution | 0.1℃ | |||
| Power | Output power | Single channel continuous working power 1800W | ||
| Stability | ±0.1%FS | |||
| 5 | Communication | LAN | ||
| 6 | Control program | Support data protection when power off, continue testing caused of power cutoff or manually stop and revise step parameter online. | ||
| Users can set protection condition, include minimum voltage, maximum voltage, minimum current, maximum current, etc. | ||||
| Users can set auxiliary channel protection condition, including minimum voltage, maximum voltage, maximum temperature, minimum temperature. | ||||
| Cycle time | Max. 9999 | |||
| Programmable step | Max. 9999 | |||
| Nested loop | Max. 10 layers | |||
| Step time range | No limit, support h, min, s | |||
| Data record | Maximum recording time:10ms,Voltage≤0.03%FS,Current≤0.03%FS | |||
| Program feature | Each step has one or multiple exit | |||
| 7 | Data display | Cycle list | Circulation number. Charge/discharge capacity, efficiency, power and average voltage, etc. | |
| Process list | Process number, working mode, process time, capacity, power, DIR, medium voltage, cut-off voltage, cut-off current, etc. | |||
| Detail list | Record number, system time, cumulative time, voltage, current, energy, power, cell temperature, etc. | |||
| 8 | Data export | Data can be exported as EXCEL, TXT or diagram. Provide charge discharge efficiency, charge discharge curve, voltage, capacity, current-time characteristic curves, etc. | ||
| 9 | Data saving | Save as files | ||
| 10 | Curves type | X-coordinate: total time, charge discharge capacity, cycle time, voltage, current, capacity. | ||
| Y-coordinate: total voltage, current, capacity, power, cell temperature(Optional), end of charge and discharge capacity, etc. | ||||
| 11 | Software protection | Power off data protection, power resume continue testing, OVP, ODP, OCP, UCP, OTP, capacity protection, abnormal voltage trend protection, abnormal current trend protection, abnormal current fluctuation protection, abnormal voltage fluctuation protection, abnormal charging voltage increase rate protection, abnormal discharge current decrease rate protection. | ||
| Hardware protection | Power off, open circuit and short circuit protection. Reverse or wrong connection protection, OCP, OVP, OTP, fixture contact abnormal protection. | |||
| Warning function | Switch off automatically under emergency, shutdown when power off, load connect and disconnect automatically. | |||
| 12 | Barcode scan | Support barcode start and binding test data, classify batteries rapidly and accurately. Support detailed battery database management functions, connect battery information with testing data base, which is easy for management of data base system. | ||
| 13 | Sampling terminals | Four-polar sampling. Copper nose or other fixture. | ||
| 14 | Energy consumption | Discharge power feedback to grid | ||
| 15 | Cooling method | Forced Fan-cooled | ||
| Noise | Less than 70dB | |||
| IP grade | IP20 | |||
| 16 | Repair | Modularized structure. Can be replaced or repair with spare parts. | ||
Product Specification
| Cabinet size | 600*800*1980(Device direct line out) |
| 600*990*1980(Device with adapter) | |
| Weight | ≤300KG |
| Channels number | 16CH |
Working Environment
Power Supply (Single Unit): AC 380V ±10%, 50 Hz ±5%.
Power Supply (Complete Cabinet): AC 380V ±10%, 50 Hz ±5%.
Maximum Input Power (Complete Cabinet): 38.4 kW (16 channels).
Ambient Temperature: 0–45°C.
Ambient Humidity: ≤ 85% RH (non-condensing).
Operating Environment: Free from strong vibration, corrosive gases, and flammable or explosive gases.
Product Display
![]()